Description
The Agilent 1290 Infinity Autosampler (Model G4226A) is a high-performance UHPLC sample injection module designed for ultrafast, high-pressure liquid chromatography separations. Operating at system pressures up to 1200 bar, it delivers ultra-low carryover, precise injection volumes, and rapid injection cycle times. Featuring a flow-through design with dynamic needle wash capabilities and flexible sample capacity (microplates, vial plates, or standard 2 mL vials), the G4226A integrates seamlessly into Agilent 1260 and 1290 Infinity LC stacks.
Features:
-
1200 Bar High-Pressure Operation: Supports high-pressure UHPLC applications without compromising injection precision or system durability.
-
Flow-Through Injection Architecture: Minimizes internal dead volume and extra-column peak broadening for sharp chromatographic peaks.
-
Ultra-Low Carryover: Flow-through design paired with active exterior needle washing drastically reduces sample contamination between runs.
-
Broad Injection Volume Range: Supports injection volumes from 0.1 uL up to 40 uL (expandable up to 900 uL with extension loops).
-
Versatile Sample Capacity: Holds up to 2 standard microplates/deep-well plates, or up to 108 standard 2 mL sample vials.
-
Temperature Control Compatible: Fully compatible with the optional Agilent 1290 Infinity Thermostat (G1330B) for sample cooling down to 4°C.
Specifications:
- Maximum Operating Pressure: 1200 bar (17,400 psi)
- Injection Range: 0.1 uL to 40 uL (standard) / up to 900 uL (extended)
- Injection Precision: < 0.25% RSD (for 5 uL to 40 uL injection volumes)
- Sample Capacity: 108 x 2 mL vials (standard tray) or 2 x 96/384-well plates
- Injection Cycle Time: < 21 seconds (without wash)
- Carryover: Typically < 0.004% (with active needle wash)
- Communication Interface: CAN bus to Agilent HPLC stack / Ethernet
- Dimensions (W x D x H): 34.5 x 43.5 x 20.2 cm (13.5 x 17.0 x 8.0 in)
- Electrical Requirements: 100-240 V AC, 50/60 Hz, 300 VA
- Weight: Approx. 15.5 kg (34.2 lbs)